•Verify/correct the nominal
values which are used to
set default thresholds
•Press Tracking for
frequency tracking options
•Nominal frequency should
match your line frequency
•Nominal Current is
whatever is detected on
circuit. Set to your spec
•Press Next when finished
Wizard Setup: Nominal Values
©2006-2008 PowerCET Corporation. All rights reserved.
80
•Standard (16-440 Hz.) "rides
through” voltage changes &
tracks the fundamental
frequency even in the absence
of voltage
•Fast (16-65 Hz.) follows
frequency changes more
closely
•Internal uses the nominal
frequency when making
measurements where no ref.
signal (e.g. DC) is available
•Specify the channel used for
external sync (A or D) and the
tracking range
•Press OK when finished
•Press Next when finished
Frequency Tracking